Mobile QR Code QR CODE : The Transactions P of the Korean Institute of Electrical Engineers
The Transactions P of the Korean Institute of Electrical Engineers

Korean Journal of Air-Conditioning and Refrigeration Engineering

ISO Journal TitleTrans. P of KIEE
  • Indexed by
    Korea Citation Index(KCI)
Title Secondary Current Range Extension of Current Transformers by Using Two Different Current Transformers and Absolute Evaluation Technique
Authors 김윤형(Kim, Yoon-Hyoung) ; 한상길(Han, Sang-Gil) ; 정재갑(Jung, Jae-Kap) ; 한상옥(Han, Sang-Ok)
Page pp.72-78
ISSN 1229-800X
Keywords Current Transformer ; Current Range Extension ; Absolute Evaluation ; Ratio Error ; Phase Displacement ; CT Test System
Abstract We have developed a current range extension method to obtain the ratio error and phase displacement of a current transformer (CT) by using absolute evaluation method and two different CTs. The method was applied to CTs under test with the current ratios in the range of 5,000 A / 1 A - 20,000 A / 1 A. The ratio error and phase displacement of the CT under test obtained in this study are consistent with those measured at the national institute in Germany using the same CT under test within an expanded uncertainty (k = 2) in the overall current ratios.